The WDS has been designed to provide combined capabilities of a bedpan ⁄ urine bottle washer disinfector,
and a utensil ⁄ bowl washer disinfector, without the additional cost and space necessitated by having separate machines.
This allows more space and budget to be allocated to more income producing activities, such as additional beds, operating theatres etc.
The WDS has gone through a thorough and exhaustive R & D program, including site trials at a number of facilities.
Malmet has listened to the needs of both clinical and engineering staff and has designed the WDS to meet the requirements of all our clients,
whether it be a large metropolitan hospital, a smaller regional hospital, or an aged care facility.
ARTG Registration Number 232450 Class IIb
— Electrical Safety Certificate: CS 10462N — Watermark Certificate: WMKA21156
Electrical Rating Model WDS Single Phase |
Volts |
240V |
| Phase ⁄ Hz |
1 ph ⁄ 50 Hz |
| Amps |
20 Amps |
Internal circuit breakers Single Phase |
Heater power supply pcb |
1 pole 20A 240Vac |
| Control power supply pcb |
1 pole 10A 240Vac |
Heating Elements Single Phase |
Steam Generator |
4500W 240V 3 x 6.25A Star |
| Water tank |
4500W 240V 3 x 6.25A Star |
| |
Electrical Rating Model WDS Three Phase |
Volts |
415V |
| Phase ⁄ Hz |
3 ph ⁄ 50 Hz |
| Amps |
20 Amps |
Internal circuit breakers Three Phase |
Heater power supply pcb |
3 pole 20A 415Vac |
| Control power supply pcb |
1 pole 10A 240Vac |
Heating Elements Three Phase |
Steam Generator |
4500W 240V 3 x 6.25A Star |
| Water tank |
7200W 240V 3 x 10A Star |
ContactorsHeating Elements Three Phase unit |
Steam heating element |
3 pole 7.5kw |
| W ⁄ tank heating element |
3 pole 7.5kw |
| ELECTRICAL Common components |
| Element over temperature cut–out |
Steam Generator |
Capillary bulb thermal cut–out Manual reset 115°C 25A |
| Water tank |
Capillary bulb thermal cut–out Manual reset 100°C 25A |
| Fuse: Transformer |
Cylinder type 5 x 20mm |
T 5A H 250Vac Time Delay |
| Processor module battery |
Coin Cell |
CR2032 Lithium 3 Volts |
| |
|
|
| |
Relative Humidity |
+30% to 70% |
| PCB HARDWARE |
| PCB (Printed Circuit Boards) |
Display LCD |
D134411 |
| Heating switch module |
D132213 |
| Control module |
D134412 |
| Processor module |
D133447 |
| Conductivity sensor |
D134418 |
| Logger data download |
Via |
USB port |
| Software loading |
Via |
USB port |
| LOADING CAPACITY |
| Wash Chamber Capacity |
Bedpans ⁄ Urine Bottles |
4 x Urinal Bottles 2 x STD Bed Pan & Lid 2 x Small slipper pans & Lid 2 x Large Slipper Pan & Lid |
| Urine Bottles only |
4 x Urine Bottles |
| Bowls ⁄ Utensils |
2 x Large Bowls 3 x Medium Bowls Smaller Bowls and Kidney Dishes in Accessory Basket 6 x Large Kidney Dishes 6 x Medium Kidney Dishes |
| WASH SYSTEM |
| Nozzles and Sprays |
Flush Nozzles |
2 x Rotary Nylon |
| Tip Sprays |
2 x WL1/2 120° 1/8 bsp ss |
| Top Sprays |
4 x WL1/4 90° 1/8 bsp ss |
| Bottom Sprays |
4 x WL1/2 60° 1/8 bsp ss |
| Urine Bottle Sprays |
4 x WL1/4 90° 1/8 bsp ss |
| Cool Down Sprays |
2 x LW1 120° 1/4 bsp ss |
| Wash Pump |
Horizontal Multistage |
750W 240V 1ph 4.6A 116.7L ⁄ Mins s ⁄ s impeller and housing |
| |
| Detergent |
Malmet specific (5Lt) |
Detergent Concentrate Caustic Alkaline (ARTGC Class 1) |
| Detergent Pump |
Self priming peristaltic Chemical feed pump |
72 rpm 240V 50Hz |
| PLUMBING |
Soil line Connection Complies to AS 2887 & AS ⁄ NZS 3500.2:2003 |
Type |
‘S’ or ‘P’ Trap (6mm PE) 80° |
| Size |
100mm ID |
Backflow Prevention AS 2845.2 – 1996 |
Air Gap (RAG) |
Water storage tank |
| Cold water inlet Supply |
Temp – Pressure (Flow) |
15 – 25°C @ 100 – 350kpa |
| Solenoid Valve 1 in 2 out |
240V – GB¾ male, WRAS Certified |
| Hot Water inlet supply |
Temp – Pressure (Flow) |
55 – 60°C @ 100 – 350kpa |
| Solenoid Valve 1 in 2 out |
240V – GB¾ male, WRAS Certified |
| Minimum water flow @ 100 Kpa |
|
9.377 L ⁄ min |
| Hose – valve to Water Tank |
Stainless Steel Braided |
10mm, WRAS Certified |
| Hoses – Mains Water Supply to Inlet Valve |
Stainless Steel Braided Assembly H ⁄ C Water |
G¾ Hex Nut Each End 1.5M x 10mm, Cert to AS ⁄ NZS 3499-2006 |
| CONSTRUCTION MATERIALS |
| Materials |
Frame |
25 x 25 x 1.6 tube SS |
| Wash chamber |
316 SS grade 4 |
| Door |
316 SS grade 4 |
| External panels |
304 SS grade 4 |
| Water tank |
316 SS grade 2B |
| Steam generator |
316 SS grade 2B |
| Fasteners |
304 ⁄ 316 SS |
| Thermal insulation chamber and water tank |
TBA |
| ENVIROMENTAL |
| Environment conditions Transport and Storage |
Fragile, Keep away from rain |
Do not stack |
| Temperature |
-5°C to +50°C |
Environment Operating Conditions |
Temperature |
+10°C to +25°C |
| Relative Humidity |
+30% to 75% |
| Shipping Details |
Height x Width x Depth (mm) |
1770 x 800 x 750 |
| Weight = Packaging + Det |
163 Kg |